Analytical Balance|Simple Notes for Students

A close-up of an analytical balance with its glass draft shield doors open: a white weighing boat holding a pile of fine white powder sits on the round steel pan, tweezers hold a small shiny calibration weight beside it, and a green spirit level bubble is visible on the base
Cover image. The whole lesson in one view: a weighing boat of powder on the steel pan inside the open draft shield, a calibration weight in tweezers, and the level bubble that keeps the balance flat.

Level: GCSE and A-level practical work, and first-year university laboratory work.

Hold a weighing boat between your fingers for a few seconds, put it on the pan of an analytical balance, and the reading will be wrong. Not wrong by a gram — wrong by a tenth of a milligram or so, the oil from your skin adding a mass the balance can see. That is the scale this instrument lives on: it notices a fingerprint. Everything in this lesson follows from that one fact. If you understand what the balance actually measures, and the dozen small things that can fool it, you can weigh anything in a school or university laboratory with confidence. If you do not, you can follow the steps perfectly and still record a number that is wrong.

What the balance actually is

An analytical balance is the laboratory balance for small masses weighed very precisely. Two numbers define it, and they are the first thing to learn. Readability is the smallest difference the display can show: for a standard analytical balance, 0.0001 g — one tenth of a milligram, four decimal places. Capacity is the largest mass it can weigh: commonly around 220 g. So the instrument weighs things up to about the mass of a small apple, and it can tell the difference between two samples that differ by a tenth of a milligram. A kitchen scale, by comparison, typically reads to 1 g; a top-pan laboratory balance to 0.01 g or 0.1 g. The analytical balance is roughly ten thousand times more sensitive than the kitchen scale — which is why it lives inside a glass box, on a heavy table, in a quiet corner, and why it demands a routine rather than a quick press of a button.

The parts of an analytical balance: the glass draft shield with its sliding doors, the weighing pan inside, the digital display, the circular level bubble, the adjustable levelling feet underneath, and the tare and calibration buttons
Figure 1. The parts that matter. The draft shield keeps air away from the pan; the level bubble and the feet keep the balance flat; the tare button zeroes a container; the calibration button checks the balance against a known mass. Nothing on this instrument is decoration.

Three parts do most of the work. The draft shield is the glass enclosure around the pan: it blocks air currents, including your own breathing, and keeps dust off the pan. The level indicator is a small circular spirit level, a bubble in a circle on the top of the balance; the levelling feet underneath are screwed in or out until the bubble sits in the centre. And the weighing pan is a small, flat, usually stainless-steel platform — small because the instrument is built for small samples, and steel because it must not corrode when a drop of something lands on it.

How it weighs: the force that holds the pan still

A modern analytical balance does not weigh the way a kitchen scale does, and it does not weigh the way the old laboratory beam balance did. Inside it is an electromagnetic force restoration system — EMFR for short — and the idea is simple once it is pictured. The pan hangs from a lever. When you put a sample on the pan, the lever tips down a fraction of a millimetre. A position sensor notices the movement at once and sends more electric current through a coil sitting inside a permanent magnet. The coil pushes the lever back up until the pan is level again. The current needed to hold the pan level is proportional to the mass on it, so the balance measures the current, converts it into grams, and shows the result. The sample never moves a weight, and no weight is ever compared against the sample: the balance measures a force, electrically, and displays it as a mass.

How electromagnetic force restoration works: a sample on the pan tips the lever down, a position sensor detects the movement, current through a coil inside a magnet pushes the lever back level, and the current needed is converted into a mass reading; beside it the old two-pan beam balance, which compared the sample against calibrated weights
Figure 2. Two ways to weigh. Left: the modern analytical balance — the coil's current holds the pan level, and the current is the measurement. Right: the old two-pan beam balance — the sample was balanced against calibrated weights on the other pan. The beam method is not how a modern balance works.

That distinction matters for two reasons. First, because the balance measures a force (a weight) and displays it as a mass, it is calibrated for the local pull of gravity: move it to a different city, or even a different floor of a tall building, and it must be levelled and recalibrated before it can be trusted. Second, because the measurement is electrical, anything that pushes or pulls on the pan — a draught, a vibration, warm air rising, static charge — is measured too, exactly as if it were mass. The balance cannot tell a real gram from a draught pretending to be one. That is the whole story of the error table later in this lesson.

A little history, because the instrument's name makes sense once it is told. Precise balances for chemical work go back to the 1750s, when Joseph Black weighed the gases given off in his experiments on "fixed air" — what we now call carbon dioxide — and needed a balance that could detect very small changes in mass. The single-pan substitution balance, the design that made precise weighing fast enough for routine laboratory work, was introduced by Erhard Mettler in 1945. The electronic EMFR balances in laboratories today are its direct descendants: one pan, no loose weights to handle, and a digital display instead of a scale and a pointer.

Four pairs of words that get mixed up

Four confusions cause most of the lost marks on this topic, so they are settled here, before the routine.

1. Readability vs capacity. Readability is the smallest difference the display can show — 0.0001 g for a standard analytical balance. Capacity is the largest mass it can weigh — commonly around 220 g. A balance can have a fine readability and a modest capacity at the same time; the two numbers describe different things. Confusing them is how students end up trying to weigh 500 g of sand on an instrument that stops at 220 g and damages its sensor.

2. Precision vs accuracy. Precision is repeatability: weigh the same thing ten times and get nearly the same reading every time. Accuracy is closeness to the true value: the reading is near the mass the sample really has. A balance can be precise but inaccurate — it gives 10.0001 g, 10.0000 g, 10.0001 g for a 10.0000 g check weight, then reads a 10.5000 g check weight as 10.4995 g. That balance needs calibration, not a steadier hand. And it can be accurate on average but imprecise, scattering readings around the true value. Good weighing needs both.

3. Calibration vs taring. Taring zeroes the display with a container on the pan, so that only the sample added afterwards is measured. It says nothing about whether the balance is right. Calibration checks — and if needed adjusts — the balance's accuracy against a known mass, either a built-in internal weight or a certified external weight. Tare before every weighing; calibrate on the laboratory's schedule, and always after the balance has been moved or levelled.

4. Repeatability vs reproducibility. Repeatability is the same person, the same balance, the same sample, one after another — how tightly the readings agree. Reproducibility is different people, or different balances, or different days — how tightly their results agree with each other. An exam that asks about "reproducibility" is asking about the second one, and the answer usually involves the balance being calibrated, the method being written down, and the sample being handled the same way by everyone.

Which balance for which job

Not every weighing needs an analytical balance, and using one when a coarser balance would do wastes time and wears out a delicate instrument. The table is the whole decision in one place.

BalanceReadabilityTypical capacityNeeds electricity?Use it for
Analytical balance0.0001 g (0.1 mg)~220 gYesReagents for standard solutions, small samples, quantitative analysis
Top-pan (precision) balance0.01 g or 0.1 gSeveral kgYesGeneral weighing, larger masses, quick work
Triple-beam balance0.1 g~610 gNoTeaching, field work, places with no power
Microbalance0.001 mg (1 µg)A few gramsYesMicrogram samples — a different instrument, not an analytical balance

The last row is there because microgram and even nanogram measurements are sometimes put on the analytical balance's account. They belong to the microbalance and the ultramicrobalance, instruments with a readability a hundred and a hundred thousand times finer, and a capacity of only a few grams. Knowing which instrument does what is exactly the kind of thing an exam question checks: "a student needs to weigh 0.0500 g of a reagent to four decimal places — which balance?" is answered by the table, not by guessing.

The mechanical member of the family, the triple-beam balance, has its own lesson on this site — Mastering Precision: Understanding the Triple Beam Balance — and the newer developments, internal calibration, touchscreens, ionisers and connectivity, are in Exploring Technological Advancements in Analytical Balances. This lesson stays with the standard analytical balance itself: the instrument, the routine, and the errors.

The weighing routine, step by step

There is a right order to weighing, and the order is the method. Learn it as a sequence, not as a list of tips.

1. Level the balance. Look at the bubble: if it is not in the centre of the circle, turn the levelling feet until it is. A balance that is not level reads low, because only part of the sample's weight acts along the direction the sensor measures.

2. Let it warm up. A balance that has just been switched on, or moved, needs time for its internal temperature to settle — manufacturers commonly recommend at least 30 minutes. That is why many laboratories leave their analytical balances switched on, in standby, all the time.

3. Zero the display. With the pan empty and the draft shield doors closed, press zero (or tare) and wait for the stability indicator — the small symbol on the display that says the reading has settled — before going on.

4. Tare the container. Put the weighing boat, watch glass or weighing bottle on the pan, close the doors, wait for the reading to settle, and press tare. The display reads 0.0000 g. Never place a chemical directly on the pan: it can corrode the steel, contaminate the next user's sample, and it is almost impossible to get every grain off again.

5. Add the sample with a spatula. Open one door, add the sample a little at a time, close the door. Add in small amounts: it is easy to add more, and very hard to take a tenth of a milligram back out of a boat.

6. Wait, then read. Wait for the stability indicator, read the display, and write the number down at once, with all four decimal places. Do not leave the sample sitting on the pan while you do something else — a hygroscopic powder will be gaining mass from the air the whole time.

7. Clean up. Remove the container, brush any spilled grains off the pan with the balance brush (never blow on it — your breath is a draught and moist air), and close the doors.

Weighing by difference — the technique for awkward powders

Some samples cannot be weighed by taring and adding, because the act of weighing them changes them. A hygroscopic powder absorbs water from the air and gains mass while it sits on the pan; a volatile solid loses mass; a powder that must be transferred completely into a flask will always leave a few grains behind in the boat. For these, the method is weighing by difference: weigh the container with the sample in it, transfer the sample, and weigh the container again. The mass that reached the flask is the first reading minus the second — whatever stayed behind in the boat is automatically accounted for, because it is still in the boat at the second weighing.

Weighing by difference in three steps: first weigh the boat with the powder in it, then tip the powder into the flask, then weigh the empty boat again; the mass that reached the flask is the first reading minus the second
Figure 4. Weighing by difference. The powder that stays behind in the boat is weighed at the second step, so it is never counted as transferred. For a hygroscopic powder, both weighings must be done quickly, before the powder has time to absorb water from the air.

Worked example 1 — weighing by difference (practice figures)

Practice figures, not measurements from a real laboratory. A weighing boat containing sodium chloride weighs 12.8456 g. The salt is tipped into a flask, and the empty boat is weighed again: 12.3456 g. What mass of salt reached the flask?

Subtract the second reading from the first: 12.8456 g − 12.3456 g = 0.5000 g. Exactly half a gram reached the flask. Notice what the method gives you for free: if 0.0010 g of salt had stayed behind in the boat, it would be part of the 12.3456 g at the second weighing, and the subtraction would still give the mass that actually left the boat. Weighing by difference does not assume a perfect transfer. It measures the transfer.

Worked example 2 — why small samples carry a big percentage error (practice figures)

Practice figures, not measurements from a real laboratory. The balance reads to ±0.0001 g. A student weighs 0.0200 g of a reagent. What is the percentage uncertainty in that mass? And what if the same balance is used to weigh 2.0000 g?

Percentage uncertainty is the uncertainty divided by the mass, times one hundred. For the small sample: 0.0001 ÷ 0.0200 × 100 = 0.5%. For the large sample: 0.0001 ÷ 2.0000 × 100 = 0.005%. The same balance, the same readability — but the small weighing is a hundred times less accurate in percentage terms. That is the practical meaning of "minimum sample weight": below a certain mass, the balance's own readability becomes a large fraction of what you are weighing. The fix is to weigh a larger amount and dilute it, or to use a more sensitive balance — not to write down more decimal places and hope.

Worked example 3 — the warm crucible that reads light (practice figures)

Practice figures, not measurements from a real laboratory. A crucible that truly weighs 25.0000 g is put on the pan while it is still warm from the oven. It reads 24.9960 g. How far light is the reading, and why?

The reading is 25.0000 g − 24.9960 g = 0.0040 g light — forty times the balance's readability, and far too large to ignore. The cause is convection: warm air rises above the hot crucible, and the rising air lifts the pan, so the balance measures less force than the crucible's weight. The remedy is the simplest in the whole lesson — let the crucible cool to room temperature, in a desiccator if it must stay dry, before it goes anywhere near the pan. A cold sample has the opposite problem: moisture from the air condenses on its cold surface, and it reads heavy until it warms up and the water evaporates.

What fools the balance

Every error on an analytical balance is a force pretending to be mass. The table gives each one its mechanism and its fix, because an exam question that names an error usually wants the mechanism, and a practical that produces a drifting reading wants the fix.

Pressure (what goes wrong)Mechanism (why it happens)Response (what to do)
Draughts and air currentsMoving air pushes on the panClose the draft shield doors; keep the balance away from windows, doors and vents; do not walk past it while weighing
VibrationThe bench shakes the panHeavy, solid table; away from centrifuges, doors and corridors
Warm sampleRising warm air lifts the pan — reads lightCool to room temperature first, in a desiccator if it must stay dry
Cold sampleMoisture condenses on the cold surface — reads heavyLet it reach room temperature before weighing
Static electricityCharge on plastic containers pushes or pulls the panUse glass or metal containers; use an ioniser; raise the humidity
Hygroscopic sampleThe powder absorbs water from the air — reading creeps upWeigh quickly, in a closed weighing bottle; weigh by difference
Balance not levelOnly part of the weight acts along the sensor's direction — reads lowCentre the bubble with the levelling feet
Fingerprints on pan or weightsSkin oils add massHandle containers with forceps or gloves; never touch the pan with bare hands
OverloadMass above the capacity strains the sensorRemove the sample at once; never weigh above the capacity
What fools the balance: a draught blowing across the open draft shield, a vibration from a nearby centrifuge shaking the bench, warm air rising from a hot crucible and lifting the pan, and static charge crackling around a plastic weighing boat
Figure 3. Four forces pretending to be mass. A draught, a vibration, warm air rising and static charge all push or pull on the pan, and the balance measures them exactly as if they were the sample. The draft shield stops only the first of the four.

Seven mistakes that sound like good lab practice

1. "Tare and calibration are the same thing." They are not. Tare zeroes a container; calibration checks accuracy against a known mass. Pressing tare on an uncalibrated balance gives you a precisely wrong number.

2. "Weighing a hot sample saves time." It wastes the weighing. A warm sample reads light by far more than the balance's readability, and the reading keeps changing as it cools.

3. "Blowing dust off the pan is fine." It is not. Your breath is a warm, moist draught: it pushes the pan and leaves moisture behind. Use the balance brush.

4. "More decimal places means more accuracy." It does not. The display shows four decimal places because the balance reads to 0.0001 g; writing down all four does not make a careless weighing accurate, and weighing 0.0200 g on it still carries a 0.5% uncertainty.

5. "A plastic weighing boat is as good as a glass one." In dry air it is not. Plastic builds up static charge, and the charge pushes and pulls on the pan. Glass or metal boats are the safe choice for precise work.

6. "If the reading is steady, it must be right." A steady reading can be steadily wrong — an unlevelled balance, an uncalibrated balance and a sample absorbing water can all give a rock-steady number that is not the true mass. Steadiness is precision, not accuracy.

7. "The balance can weigh anything up to its capacity." It can weigh it without damage, but the result may still be useless: a 200 g sample weighed to 0.0001 g is fine, while a 0.0200 g sample weighed on the same balance carries a 0.5% uncertainty. Capacity is a limit, not a promise.

How to answer an exam question on this topic

Use the sequence instrument → principle → routine step → error → fix. Name the instrument and its readability, say how it measures, name the step of the routine that matters, name the error that step prevents, and say what the reading would do without it.

Model paragraph: An analytical balance reads to 0.0001 g and measures mass by electromagnetic force restoration: current through a coil holds the pan level, and the current needed is converted into a mass reading. Before weighing, the balance is levelled so the bubble is centred, and the sample is allowed to reach room temperature. A warm sample must not be weighed, because rising warm air lifts the pan and the reading comes out light — by 0.0040 g in the practice example, forty times the balance's readability. The sample is weighed in a tared weighing boat with the draft shield doors closed, and the reading is taken only when the stability indicator shows the display has settled. That paragraph names the instrument, the principle, two routine steps, one error with its mechanism and its size, and the fix — which is what the marks are for.

Frequently asked questions

Why does the balance need a draft shield when the laboratory has no wind?
Because "no wind" is not the same as "no air movement". Your breathing, someone walking past, warm air rising from a radiator — all of these move air across the pan, and the balance can feel them. The shield is there for the draughts you cannot see.

How often should I calibrate my balance?
As often as your laboratory's procedure says, and always after the balance has been moved, re-levelled, or put back into service. Many laboratories check the balance daily with a check weight and calibrate before critical work. There is no single interval, such as "monthly", that suits every laboratory.

Can I weigh a liquid on an analytical balance?
Yes, in a closed container — a weighing bottle or a stoppered flask — tared first. Never pour a liquid directly onto the pan, and remember that a volatile liquid loses mass while it sits there, so weigh it quickly and by difference if the loss matters.

Why is my reading drifting upward?
The most common reason is a hygroscopic sample absorbing water from the air. Other causes are a sample still warming up or cooling down, static charge, or vibration. Work through the error table: check the doors, the temperature of the sample, and whether the sample is one that absorbs water.

What is the difference between an analytical balance and a precision (top-pan) balance?
Readability and capacity. The analytical balance reads to 0.0001 g with a capacity around 220 g and a draft shield; the top-pan balance reads to 0.01 g or 0.1 g, weighs kilograms, and has no draft shield. Use the analytical balance when the mass is small and the accuracy matters, and the top-pan balance for everything else.

Do I really need to write down all four decimal places?
Yes. The fourth decimal place is the balance's readability, and dropping it throws away the precision you used the instrument for. Write 0.5000 g, not 0.5 g — the trailing zeros are information, not decoration.

Continue learning

Sources and further reading

These notes were written from the sources below and from the site's own linked lessons. The counting figures in the worked examples are invented for teaching and are labelled as such.

  • OpenStax, Chemistry 2e, Chapter 1: Essential Ideas — measurement, accuracy and precision, significant figures (2019).
  • LibreTexts, Analytical Chemistry 2.0 (Harvey) — the analytical balance, weighing by difference, and sources of error in weighing.
  • Mettler Toledo, Good Weighing Practice (GWP) — calibration, minimum sample weight, and routine testing of balances.
  • Sartorius, weighing guides — electromagnetic force restoration and the construction of the analytical balance.
  • Royal Society of Chemistry, practical chemistry guidance — safe weighing of hazardous chemicals and the use of fume cupboards.
  • PreachBio, Mastering the Art of Laboratory Glassware — weighing boats, weighing bottles and desiccators, for the equipment in this lesson.
  • PreachBio, Mastering Precision: Understanding the Triple Beam Balance — the mechanical balance, for the comparison table.

20 Explained MCQs — answers hidden until opened

Open each question to reveal its answer and explanation. The interactive quiz below uses a different set of questions.

1. What does a readability of 0.1 mg mean on an analytical balance?

A. The largest mass the balance can hold is 0.1 mg   B. The balance must be calibrated every 0.1 mg   C. The balance weighs a sample in 0.1 mg steps only when it is warm   D. The smallest mass difference the display can show is 0.0001 g
Answer: D. Readability is the smallest difference the display can show: 0.1 mg is 0.0001 g, four decimal places. It says nothing about the largest mass the balance can hold — that is the capacity.

2. What is the typical capacity of a standard analytical balance?

A. About 2,000 g   B. About 220 g   C. About 20 kg   D. About 2 g
Answer: B. A standard analytical balance commonly has a capacity around 220 g — about the mass of a small apple. Kilograms belong to top-pan balances; 2 g is far below what it can hold.

3. How does a modern analytical balance actually measure mass?

A. It counts the vibrations of the pan   B. A spring stretches under the sample   C. Electromagnetic force restoration: current through a coil holds the pan level   D. It compares the sample against calibrated weights on the other pan
Answer: C. A modern balance uses electromagnetic force restoration (EMFR): a coil inside a magnet generates the force that holds the pan level, and the current needed is converted into a mass reading. Comparing against weights is how the old two-pan beam worked.

4. In electromagnetic force restoration, what is directly measured?

A. The electric current needed to hold the pan level   B. The time the reading takes to settle   C. The volume of the sample   D. The temperature of the coil
Answer: A. The sensor detects the pan tipping down, more current is sent through the coil to push it back level, and that current — proportional to the mass — is what the balance measures and converts into grams.

5. What is the draft shield for?

A. To keep the sample warm   B. To store the calibration weights   C. To block air movements, breezes and dust from the pan   D. To display the reading
Answer: C. The glass enclosure blocks air currents — including your own breathing — and dust, both of which can change a reading at the 0.1 mg level. It does not warm the sample or store anything.

6. What does the level bubble tell you?

A. Whether the balance is calibrated   B. Whether the balance is level   C. Whether the sample is dry   D. Whether the draft shield doors are shut
Answer: B. The circular spirit level shows whether the balance is flat: the bubble must sit in the centre of the circle. If it does not, turn the levelling feet until it does.

7. What happens to the reading if the balance is not level?

A. It reads high   B. It reads low   C. The display goes blank   D. It does not change at all
Answer: B. When the balance is tilted, only part of the sample's weight acts along the direction the sensor measures, so the reading comes out low. Level the balance before you trust it.

8. What does the tare button do?

A. It zeroes the display with a container on the pan, so only the sample is measured   B. It calibrates the balance   C. It switches the display off   D. It locks the draft shield doors
Answer: A. Taring zeroes out the container's mass: put the weighing boat on the pan, press tare, and the display reads 0.0000 g, so everything added afterwards is the sample alone.

9. What is the difference between taring and calibrating?

A. Taring zeroes a container's mass; calibrating checks and adjusts accuracy against a known mass   B. Taring is done once a year; calibrating is done before every weighing   C. Calibrating zeroes the container; taring checks the accuracy   D. They are the same thing, with two names
Answer: A. Tare removes the container from the measurement; calibration checks the balance itself against a known mass, using a built-in internal weight or a certified external weight. Tare before every weighing; calibrate on the laboratory's schedule.

10. Which sample should you weigh by difference?

A. A warm crucible   B. A liquid in a sealed bottle   C. A hygroscopic powder that gains mass while it sits in the air   D. A large metal block
Answer: C. A hygroscopic powder absorbs water from the air and gains mass while it sits on the pan, so you weigh the boat with the powder, transfer it, and weigh the boat again. The subtraction gives the mass that actually left the boat.

11. Practice figures, not real measurements: a weighing boat containing sodium chloride weighs 12.8456 g. The salt is tipped into a flask, and the empty boat weighs 12.3456 g. What mass of salt reached the flask?

A. 12.3456 g   B. 25.1912 g   C. 0.0500 g   D. 0.5000 g
Answer: D. Subtract the second reading from the first: 12.8456 g − 12.3456 g = 0.5000 g. Whatever stayed behind in the boat is part of the second weighing, so it is never counted as transferred.

12. Practice figures: the balance reads to ±0.0001 g. A student weighs 0.0200 g of a reagent. What is the percentage uncertainty in that mass?

A. 5%   B. 0.5%   C. 0.05%   D. 0.005%
Answer: B. Percentage uncertainty is the uncertainty divided by the mass, times one hundred: 0.0001 ÷ 0.0200 × 100 = 0.5%. On a 0.0200 g sample the balance's own readability is half a percent of the mass.

13. Practice figures: the same balance, readability ±0.0001 g, is used to weigh 2.0000 g. What is the percentage uncertainty now?

A. 5%   B. 0.05%   C. 0.5%   D. 0.005%
Answer: D. 0.0001 ÷ 2.0000 × 100 = 0.005%. The same balance is a hundred times more accurate in percentage terms on the larger sample — which is why weighing too small a sample is a real error, not a rounding choice.

14. Why does a warm sample read light on an analytical balance?

A. Rising warm air lifts the pan, so the balance measures less force   B. The draft shield melts and lets air in   C. Heat makes the sample lose mass   D. The balance switches itself off when it is warm
Answer: A. Warm air rises above the hot sample and lifts the pan — a convection current pushing up. The balance measures less force than the sample's weight, so the reading comes out light. Cool the sample to room temperature first.

15. A cold sample taken from the fridge reads heavy. Why?

A. Cold makes mass increase   B. Moisture from the air condenses on the cold surface   C. The balance freezes   D. Static electricity increases when things are cold
Answer: B. Water vapour from the laboratory air condenses on the cold surface, adding real mass to the sample until it warms up and the water evaporates. Let a cold sample reach room temperature before weighing it.

16. Why should you never place a chemical directly on the weighing pan?

A. It can corrode the pan and contaminate the next sample weighed on it   B. It makes the display brighter   C. It is always too heavy for the pan   D. It will melt the pan
Answer: A. Chemicals can corrode the steel pan and leave traces that contaminate the next user's sample, and loose grains are almost impossible to remove completely. Use a weighing boat, watch glass or weighing bottle.

17. Static electricity makes the reading jump. Which container makes static worst?

A. A plastic weighing boat in dry air   B. A metal weighing boat   C. A glass weighing bottle   D. A porcelain crucible
Answer: A. Plastic builds up static charge, especially in dry air, and the charge pushes and pulls on the pan. Glass and metal do not hold charge the same way, so they are the safe choice for precise work.

18. What is the difference between precision and accuracy?

A. Precision is about capacity; accuracy is about speed   B. Accuracy is about repeatability; precision is about the true value   C. They mean the same thing   D. Precision is repeatability; accuracy is closeness to the true value
Answer: D. Precision means the readings agree with each other when you repeat the weighing; accuracy means the reading is close to the true value. A balance can be precise but inaccurate — steady readings that are all wrong — which is why calibration matters.

19. A balance gives 10.0001 g, 10.0000 g and 10.0001 g for a 10.0000 g check weight, but reads a 10.5000 g check weight as 10.4995 g. The balance is…

A. Broken beyond repair   B. Neither precise nor accurate   C. Accurate but imprecise   D. Precise but inaccurate
Answer: D. The readings repeat tightly, so the balance is precise; but it is 0.0005 g out on the heavier weight, so it is inaccurate and needs calibration. It is not broken — a calibration against a known mass corrects it.

20. Which claim stays inside the evidence of the practice figures in this lesson?

A. In the practice figures, a 0.0200 g sample carried a 0.5% uncertainty while a 2.0000 g sample carried 0.005%   B. All warm samples read exactly 0.0040 g light   C. Every analytical balance reads to 0.1 mg   D. Every balance has a capacity of 220 g
Answer: A. Only the claim that repeats the practice figures is safe: on this balance, 0.0001 ÷ 0.0200 × 100 = 0.5% and 0.0001 ÷ 2.0000 × 100 = 0.005%. The others turn one worked example into a rule about all balances and all samples.

Separate Interactive Quiz — 20 different questions

These quiz questions are different from the MCQs above. Select an answer for each question. You will not see whether an answer is correct until you submit all 20 questions. Use Reset to try again.